DiamondTomeTM Facial Wands
The Heart of Particle Free MicroExfoliation

There are 4 DiamondTomeTM facial wands to choose from. The facial wands are available in two grits and two different tip configurations, round and flat.

Fine & Medium Facial Wands

The Fine and Medium wands are used for exfoliation of the stratum corneum (dead skin cells). This is ideal for the individual who has healthy skin and wants to look radiant and beautiful the evening of treatment. With the top layer of dead tissue gone, light is reflected and the skin looks luminous. It is truly the "lunch time peel."

Round & Flat Lymphatic Massage Wands

These wands are used for lymphatic drainage.
DiamondTome™ Wands Cleaning and Sterilization

The wand is a reusable device. It is considered a semi-critical device, which comes into contact with intact skin. High levels of disinfecting are the minimum process requirement.

Immediately after each use, the wand should be cleaned by soaking in enzymatic detergent for 15 minutes. The wand should then be washed in hot running water and the abrasive surface scrubbed with a brush to remove protein material.

The wand then can be placed in a disinfecting solution, Read the label on the solution for the proper soak time.

Caution: Carefully follow the labeling instructions of the disinfectant. Dilution and the soak times are important, the common soak time is 10 minutes. Soaking beyond the recommended times will damage and corrode the wands and the diamond surface will be ruined.

Steam Sterilization The wand can withstand steam sterilization. Steam sterilization should always be used if the surface of the skin may be penetrated.

Gravity Displacement Steam Sterilization:
If the wand is sterilized in a unwrapped tray the "flash" cycle may be used. This is five minutes at a minimum temperature of 132ºC (270ºF) in a steam autoclave.